Nojiri C, Kijima T, Maekawa J, Horiuchi K, Kido T, Sugiyama T, Mori T, Sugiura N, Asada T, Shimane H, Ozaki T, Suzuki M, Akamatsu T, Akutsu T
Terumo Corporation, R&D Center, Kanagawa, Japan.
ASAIO J. 1999 May-Jun;45(3):199-203. doi: 10.1097/00002480-199905000-00018.
The research group of the Terumo Corporation, the NTN Corporation, and Setsunan University (T. Akamatsu) has been developing an implantable left ventricular assist system (ILVAS) featuring a centrifugal blood pump with a magnetically suspended impeller (MSCP). The impeller of the MSCP is suspended by a magnetic bearing, providing contact-free rotation of the impeller inside the pump housing. Thus the MSCP is expected to provide years of long-term durability. Ex vivo chronic sheep experiments using the extracorporeal model (Model I) demonstrated long-term durability, nonthrombogenicity, and a low hemolysis rate (plasma free Hb <6 mg/dl) for more than 2 years. The prototype implantable model (Model II; 196 ml, 400 g) was evaluated ex vivo in 2 sheep and intrathoracically implanted in a small sheep (45 kg). These experiments were terminated at 70, 79, and 17 days, respectively, because of blood leakage through the connector system within the housing of Model II. There was no thrombus formation on the retrieved pump surfaces. A new connector system was introduced to the Model II pump (modified Model II), and the pump was intrathoracically implanted in a sheep. Pump flow rate was maintained at 3-7 L/min at 1700-1800 rpm. The temperature elevation on the surfaces of the motor and the electromagnet inside the pump casing was kept less than 6 degrees C. The temperature of the tissue adjacent to the pump casing became normal 10 days postoperatively. The sheep survived for more than 5 months without any sign of mechanical failure or thromboembolic complication. In vitro real-time endurance tests of motor bearings made of stainless steel and silicone nitride have been conducted for more than 1 year without any sign of bearing wear. The next prototype system (Model III), with an implantable controller and a new MSCP with reduced input power, has been developed with a view toward a totally implantable LVAS.
泰尔茂株式会社、日本精工株式会社和摄南大学(赤松泰)的研究团队一直在研发一种植入式左心室辅助系统(ILVAS),其特点是采用带有磁悬浮叶轮的离心血泵(MSCP)。MSCP的叶轮由磁轴承悬浮,使叶轮在泵壳内实现无接触旋转。因此,预计MSCP能提供数年的长期耐用性。使用体外模型(模型I)进行的体外慢性绵羊实验表明,该系统具有超过2年的长期耐用性、抗血栓形成性和低溶血率(血浆游离血红蛋白<6mg/dl)。原型植入式模型(模型II;196毫升,400克)在2只绵羊身上进行了体外评估,并植入一只小羊(45千克)的胸腔内。由于模型II泵壳内的连接器系统出现血液泄漏,这些实验分别在70天、79天和17天时终止。回收的泵表面没有血栓形成。一种新的连接器系统被引入到模型II泵(改进型模型II)中,并将该泵植入一只绵羊的胸腔内。泵的流速在1700 - 1800转/分钟时维持在3 - 7升/分钟。泵壳内电机和电磁铁表面的温度升高保持在6摄氏度以下。术后10天,与泵壳相邻的组织温度恢复正常。这只绵羊存活了5个多月,没有出现任何机械故障或血栓栓塞并发症的迹象。由不锈钢和氮化硅制成的电机轴承的体外实时耐久性测试已经进行了1年多,没有出现任何轴承磨损的迹象。下一代原型系统(模型III)已开发出来,它带有植入式控制器和输入功率降低的新型MSCP,旨在实现完全植入式左心室辅助系统。