J Cataract Refract Surg. 1997 Oct;23(8):1213-6. doi: 10.1016/s0886-3350(97)80318-5.

Abstract

PURPOSE

To compare the accuracy of portable automated keratometry (PAK) with that of manual keratometry (MK) in measuring corneal power for intraocular lens calculations.

SETTING

Wilford Hall Medical Center, San Antonio, Texas, USA.

METHODS

In Part 1 of the study, five ophthalmic technicians performed keratometric analysis in 20 eyes in 10 volunteers using both manual and automated methods to determine the relative accuracy and reproducibility of each instrument. In Part 2, both MK and PAK were prospectively performed in 11 patients having cataract surgery to compare the accuracy of each instrument in predicting postoperative refractive outcome.

RESULTS

The difference between instruments in determining the average corneal power in all eyes was less than 0.10 diopter (D) (MK = 43.84 D, PAK = 43.93 D). Portable keratometry demonstrated less variability in measurements in each eye (average standard deviation, MK = 0.30 D, PAK = 0.11 D; average range, MK = 1.08 D, PAK = 0.44 D). The mean absolute refractive error (difference between the actual refractive outcome and predicted refractive outcome) was 0.37 D +/- 0.30 (SD) using MK values and 0.45 +/- 0.19 D using PAK values.

CONCLUSIONS

Portable automated keratometry is a simple keratometric technique that appeared to be as accurate as but with less variability than manual keratometry in determining corneal power for cataract surgery.

摘要

目的

比较便携式自动角膜曲率计(PAK)与手动角膜曲率计(MK)在测量角膜屈光力以进行人工晶状体计算时的准确性。

设置

美国得克萨斯州圣安东尼奥市威尔福德·霍尔医疗中心。

方法

在研究的第1部分中,5名眼科技术人员使用手动和自动方法对10名志愿者的20只眼睛进行角膜曲率分析,以确定每种仪器的相对准确性和可重复性。在第2部分中,对11名接受白内障手术的患者前瞻性地进行了MK和PAK检查,以比较每种仪器在预测术后屈光结果方面的准确性。

结果

在所有眼睛中,两种仪器在确定平均角膜屈光力方面的差异小于0.10屈光度(D)(MK = 43.84 D,PAK = 43.93 D)。便携式角膜曲率计在每只眼睛的测量中显示出较小的变异性(平均标准差,MK = 0.30 D,PAK = 0.11 D;平均范围,MK = 1.08 D,PAK = 0.44 D)。使用MK值时,平均绝对屈光误差(实际屈光结果与预测屈光结果之间的差异)为0.37 D±0.30(标准差),使用PAK值时为0.45±0.19 D。

结论

便携式自动角膜曲率计是一种简单的角膜曲率测量技术,在为白内障手术确定角膜屈光力时,其准确性似乎与手动角膜曲率计相当,但变异性较小。