Rubin J M, Bude R O, Fowlkes J B, Spratt R S, Carson P L, Adler R S
Department of Radiology, University of Michigan Hospitals, Ann Arbor 48109, USA.
Radiology. 1997 Dec;205(3):757-65. doi: 10.1148/radiology.205.3.9393532.
To normalize the power Doppler ultrasound (US) signal to the expected signal from 100% blood in the calculation of a fractional moving blood volume estimate.
To locate the signal from flowing blood with a consistent backscatter coefficient, the authors estimated the knee of the cumulative Doppler power distribution function. They used a flow-tube phantom to test the use of this knee to locate a radial position that would fall into a region of high shear stress and minimal rouleaux formation. They also studied how well the method normalized fractional moving blood volume estimates of the right renal cortex in a volunteer when simulating different body habitus and in a group of six healthy volunteers to estimate variability.
Over five flow velocities and over undersaturated to severely oversaturated receiver gains, the calculated flow-tube area was a mean 89% +/- 7 (+/- standard deviation) of a standard. In humans, the technique normalized the fractional moving blood volume estimates over an 8-dB receiver gain variation; the mean +/- standard deviation of fractional moving blood volume estimates for the six volunteers was 37.6% +/- 3.6.
Vascularity estimates with power Doppler US are feasible with a normalization scheme based on the cumulative Doppler power distribution function.