The x-ray generator delivers the electrical power to energize the x-ray tube and permits the selection of x-ray energy, x-ray quantity, and exposure time. Major internal components of the generator include transformers, diodes and rectifier circuits, filament and stator circuits, timer switches, and kilovolt and milliampere meters. Single-phase, three-phase, high-frequency, and constant potential generators produce different voltage waveforms (ripple) and x-ray beam spectra. Phototimer and automatic brightness control subsystems measure radiation exposure incident on the image receptor to give instantaneous feedback for optimal radiographic film densities and fluoroscopic image brightness, respectively. At the generator control console, the operator sets the tube voltage, tube current, exposure time, phototimer film density, spot film acquisition, and fluoroscopic parameters. Selection of generator power and options depends on the intended clinical use. X-ray tube focal spot size and power loading capability should be matched to the x-ray generator and clinical imaging requirements. Single and multiple exposure rating charts as well as anode and housing thermal characteristic charts indicate power input and dissipation rates specific to a generator and x-ray tube target and housing.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验