Mubata C D, Bidmead A M, Ellingham L M, Thompson V, Dearnaley D P
The Joint Department of Physics, The Royal Marsden NHS Trust and The Institute of Cancer Research, London, UK.
Int J Radiat Oncol Biol Phys. 1998 Jan 1;40(1):221-31. doi: 10.1016/s0360-3016(97)00551-8.
The use of escalated radiation doses to improve local control in conformal radiotherapy of prostatic cancer is becoming the focus of many centers. There are, however, increased side effects associated with increased radiotherapy doses that are believed to be dependent on the volume of normal tissue irradiated. For this reason, accurate patient positioning, CT planning with 3D reconstruction of volumes of interest, clear definition of treatment margins and verification of treatment fields are necessary components of the quality control for these procedures. In this study electronic portal images are used to (a) evaluate the magnitude and effect of the setup errors encountered in patient positioning techniques, and (b) verify the multileaf collimator (MLC) field patterns for each of the treatment fields.
The Phase I volume, with a planning target volume (PTV) composed of the gross tumour volume (GTV) plus a 1.5 cm margin is treated conformally with a three-field plan (usually an anterior field and two lateral or oblique fields). A Phase II, with no margin around the GTV, is treated using two lateral and four oblique fields. Portal images are acquired and compared to digitally reconstructed radiographs (DRR) and/or simulator films during Phase I to assess the systematic (CT planning or simulator to treatment error) and the daily random errors. The match results from these images are used to correct for the systematic errors, if necessary, and to monitor the time trends and effectiveness of patient imobilization systems used during the Phase I treatment course. For the Phase II, portal images of an anterior and lateral field (larger than the treatment fields) matched to DRRs (or simulator images) are used to verify the isocenter position 1 week before start of Phase II. The Portal images are acquired for all the treatment fields on the first day to verify the MLC field patterns and archived for records. The final distribution of the setup errors was used to calculate modified dose-volume histograms (DVHs). This procedure was carried out on 36 prostate cancer patients, 12 with vacuum-molded (VacFix) bags for immobilization and 24 with no immobilization.
The systematic errors can be visualized and corrected for before the doses are increased above the conventional levels. The requirement for correction of these errors (e.g., 2.5 mm AP shift) was demonstrated, using DVHs, in the observed 10% increase in rectal volume receiving at least 60 Gy. The random (daily) errors observed showed the need for patient fixation devices when treating with reduced margins. The percentage of fields with displacements of < or = 5.0 mm increased from 82 to 96% with the use of VacFix bags. The rotation of the pelvis is also minimized when the bags are used, with over 95% of the fields with rotations of < or = 2.0 degrees compared to 85% without. Currently, a combination of VacFix and thermoplastic casts is being investigated.
The systematic errors can easily be identified and corrected for in the early stages of the Phase I treatment course. The time trends observed during the course of Phase I in conjunction with the isocenter verification at the start of Phase II give good prediction of the accuracy of the setup during Phase II, where visibility of identifiable structures is reduced in the small fields. The acquisition and inspection of the portal images for the small Phase I fields has been found to be an effective way of keeping a record of the MLC field patterns used. Incorporation of the distribution of the setup errors into the planning system also gives a clearer picture of how the prescribed dose was delivered. This information can be useful in dose-escalation studies in determining the relationship between the local control or morbidity rates and prescribed dose.
在前列腺癌适形放疗中,使用递增辐射剂量以改善局部控制正成为许多中心关注的焦点。然而,放疗剂量增加会带来更多副作用,据信这些副作用取决于受照射正常组织的体积。因此,精确的患者定位、利用感兴趣体积的三维重建进行CT规划、明确治疗边界的定义以及治疗野的验证是这些程序质量控制的必要组成部分。在本研究中,使用电子射野影像来(a)评估患者定位技术中所遇到的摆位误差的大小和影响,以及(b)验证每个治疗野的多叶准直器(MLC)野形状。
I期体积,其计划靶体积(PTV)由大体肿瘤体积(GTV)加上1.5厘米的边界组成,采用三野计划(通常一个前野和两个侧野或斜野)进行适形治疗。II期,GTV周围无边界,采用两个侧野和四个斜野进行治疗。在I期期间获取射野影像,并与数字重建射线照片(DRR)和/或模拟定位片进行比较,以评估系统误差(CT规划或模拟定位到治疗的误差)和每日随机误差。如果有必要,这些影像的匹配结果用于校正系统误差,并监测I期治疗过程中所使用的患者固定系统的时间趋势和有效性。对于II期,在II期开始前1周,使用与DRR(或模拟定位影像)匹配的一个前野和一个侧野(大于治疗野)的射野影像来验证等中心位置。在第一天获取所有治疗野的射野影像以验证MLC野形状,并存档记录。摆位误差的最终分布用于计算修正剂量体积直方图(DVH)。该程序在36例前列腺癌患者中进行实施,其中12例使用真空成型(VacFix)袋进行固定,24例未进行固定。
在剂量增加到高于传统水平之前,可以直观显示并校正系统误差。使用DVH证明了对这些误差进行校正的必要性(例如,前后方向2.5毫米的移位),观察到接受至少60 Gy剂量的直肠体积增加了10%。观察到的随机(每日)误差表明,在使用较小边界进行治疗时需要患者固定装置。使用VacFix袋时,移位≤5.0毫米的射野百分比从82%增加到96%。使用这些袋子时,骨盆的旋转也降至最低,旋转≤2.0度的射野超过95%,而未使用袋子时为85%。目前,正在研究VacFix和热塑性模型的联合使用。
在I期治疗过程的早期阶段,可以轻松识别并校正系统误差。在I期过程中观察到的时间趋势以及II期开始时对等中心的验证,能够很好地预测II期治疗过程中的摆位准确性,在II期小射野中可识别结构的可见性降低。已发现对I期小射野的射野影像进行采集和检查是记录所使用的MLC野形状的有效方法。将摆位误差的分布纳入计划系统,也能更清楚地了解处方剂量的给予方式。该信息在剂量递增研究中对于确定局部控制或发病率与处方剂量之间的关系可能有用。