Sittek H, Perlet C, Herrmann K, Linsmeier E, Kolem H, Untch M, Kessler M, Reiser M
Institut für Radiologische Diagnostik, Ludwig-Maximilians-Universität, München.
Radiologe. 1997 Sep;37(9):685-91. doi: 10.1007/s001170050269.
To answer the following questions: whether reliable detection of lesions is possible in low-field-MRI-system (Magnetom Open 0.2 T) equipped with currently available hard- and software components in comparison to high field system (Magnetom Impact 1.0 T). Furthermore, whether localization of lesions suspect in MR-mammography can be realized in MR system of low field (Magnetom Open 0.2 T).
In 11 patients, suspect lesions were diagnosed in diagnostic MR-mammography acquired with high field system (Magnetom Impact 1.0 T) and were compared to low field MR-mammographies of 0.2 T (Magnetom Open 0.2 T). In six of the 11 patients a suspect lesion was localized using wire marking.
All lesions considered suspect in diagnostic MR-mammography (Magnetom Impact 1.0 T) were also clearly identified in the 0.2 T system (Magnetom Open). In six cases wire marking was performed without any complications and with an accuracy of < or = 0.5 cm distance to the lesion.
Although studies in the 0.2 T system clearly showed inferior SNR (34.6 vs. 83.1) and CNR (14.6 vs. 43.5) compared to studies with the high field system, all lesions considered suspect in diagnostic MR-mammography were reliably identified also in 0.2 T studies. Due to its open construction permitting permanent access to the breast and due to sufficient image quality, the Magnetom Open is suitable for interventions on the breast.
回答以下问题:与高场系统(Magnetom Impact 1.0 T)相比,配备当前可用硬件和软件组件的低场MRI系统(Magnetom Open 0.2 T)是否能够可靠地检测病变。此外,低场(Magnetom Open 0.2 T)MR系统是否能够实现乳腺磁共振成像中可疑病变的定位。
对11例患者进行研究,在高场系统(Magnetom Impact 1.0 T)下进行乳腺磁共振诊断成像,发现可疑病变,并与0.2 T(Magnetom Open 0.2 T)的低场乳腺磁共振成像进行比较。11例患者中有6例通过金属丝标记对可疑病变进行了定位。
在乳腺磁共振诊断成像(Magnetom Impact 1.0 T)中被视为可疑的所有病变,在0.2 T系统(Magnetom Open)中也能清晰识别。6例进行了金属丝标记,未出现任何并发症,标记位置与病变的距离误差≤0.5 cm。
尽管与高场系统相比,0.2 T系统的研究明显显示出较低的信噪比(34.6对83.1)和对比噪声比(14.6对43.5),但在乳腺磁共振诊断成像中被视为可疑的所有病变在0.2 T的研究中也能被可靠识别。由于其开放式结构允许对乳房进行持续观察,且图像质量足够,Magnetom Open适用于乳房干预操作。