使用基于X射线的双能光子吸收法测量大鼠全身骨量和身体成分的精密度与准确性。

Precision and accuracy of total body bone mass and body composition measurements in the rat using x-ray-based dual photon absorptiometry.

作者信息

Makan S, Bayley H S, Webber C E

机构信息

Department of Nuclear Medicine, Hamilton Health Sciences Corporation, ON, Canada.

出版信息

Can J Physiol Pharmacol. 1997 Oct-Nov;75(10-11):1257-61.

PMID:9431451
Abstract

Lean body mass (LBM), total body bone mineral mass (BMC), total body bone areal density (BMD), and body fat mass (FM) were measured in rats by dual photon absorptiometry (DXA), using two different instruments. The coefficients of variation for repeated measurements of LBM and FM were about 0.4 and 2.5%, respectively, over an animal body weight range of 150 to 600 g. For BMC and BMD, the coefficients of variation were less than 2%. The correlation coefficients for LBM, FM, BMC, and BMD measured on the two densitometers were all greater than 0.94. The slope of the regression line relating LBM measured by DXA and LBM measured by carcass analysis was 0.999, and the correlation coefficient was 0.99. For FM the slope was 1.05, and the correlation coefficient was 0.98. BMC measures by DXA were falsely low in small animals. For larger animals, the correlation between BMC and ash weight was 0.93, but the slope of the regression line was 0.78. DXA measures of LBM and FM were accurate and reproducible for rats weighing between 150 and 600 g. There was a size-dependent error in BMC, which will be significant in longitudinal measurements of bone mass.

摘要

使用两种不同的仪器,通过双能X线吸收法(DXA)测量大鼠的瘦体重(LBM)、全身骨矿物质质量(BMC)、全身骨面积密度(BMD)和体脂质量(FM)。在150至600克的动物体重范围内,LBM和FM重复测量的变异系数分别约为0.4%和2.5%。对于BMC和BMD,变异系数小于2%。在两台密度仪上测量的LBM、FM、BMC和BMD的相关系数均大于0.94。DXA测量的LBM与胴体分析测量的LBM之间的回归线斜率为0.999,相关系数为0.99。对于FM,斜率为1.05,相关系数为0.98。DXA测量的BMC在小动物中偏低。对于较大的动物,BMC与骨灰重量之间的相关性为0.93,但回归线的斜率为0.78。DXA对体重在150至600克之间的大鼠的LBM和FM测量准确且可重复。BMC存在大小依赖性误差,这在骨量的纵向测量中会很显著。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验