Wildhaber J H, Hayden M J, Dore N D, Devadason S G, LeSouëf P N
Department of Respiratory Medicine, Princess Margaret Hospital for Children, Subiaco, Western, Australia.
Chest. 1998 Jan;113(1):186-91. doi: 10.1378/chest.113.1.186.
The aim of our study was to determine the in vitro delivery of salbutamol from a pressurized metered-dose inhaler (pMDI) containing hydrofluoroalkane (HFA) propellant through various delivery devices to four models of a pediatric lung.
To determine the effect of electrostatic charge, delivery of salbutamol was initially assessed with a multistage liquid impinger (MSLI) through an inline nonchamber device (Baxter MDI Adapter) and a small (Aerochamber MV) and a large (Nebuhaler) inline chamber device. Following this, the delivery was assessed to four lung models appropriate for a child of 70 kg, 50 kg, 15 kg, and 4 kg, with the same three reduced static devices inserted directly into a pediatric ventilator circuit.
Reduction of electrostatic charge improved small particle delivery through holding chambers to the MSLI by 12 to 14%. In the ventilator model, the mean delivery was between 1.9% and 5.4% for the nonchamber device, between 14.3% and 27.2% for the small holding chamber, and between 7.2% and 25.7% for the large holding chamber. Delivery was the least efficient in the 4-kg model compared to the 70-kg, 50-kg, and 15-kg models.
Salbutamol from an HFA pMDI is delivered efficiently through inline holding chambers with reduced static in pediatric ventilator settings. A large holding chamber has no advantage over a small holding chamber. In addition, salbutamol delivery is more efficient through a holding chamber than through a nonchamber device.
我们研究的目的是确定含有氢氟烷烃(HFA)推进剂的压力定量吸入器(pMDI)中的沙丁胺醇通过各种给药装置向四种儿科肺部模型的体外递送情况。
为了确定静电荷的影响,首先使用多级液体撞击器(MSLI)通过在线无腔装置(百特MDI适配器)以及小型(Aerochamber MV)和大型(Nebuhaler)在线腔室装置评估沙丁胺醇的递送情况。在此之后,将相同的三种减少静电的装置直接插入儿科呼吸机回路,评估向适合70kg、50kg、15kg和4kg儿童的四种肺部模型的递送情况。
静电荷的减少使通过储雾罐向MSLI的小颗粒递送提高了12%至14%。在呼吸机模型中,无腔装置的平均递送率在1.9%至5.4%之间,小型储雾罐在14.3%至27.2%之间,大型储雾罐在7.2%至25.7%之间。与70kg、50kg和15kg模型相比,4kg模型中的递送效率最低。
在儿科呼吸机设置中,HFA pMDI中的沙丁胺醇通过减少静电的在线储雾罐有效递送。大型储雾罐并不比小型储雾罐有优势。此外,通过储雾罐递送沙丁胺醇比通过无腔装置更有效。