[放射治疗设备RTA的辐射深度剂量曲线测量]

[Measurements of radiation depth-dose curves of radiotherapy device RTA].

作者信息

Vasil'ev V N, Kokontsev A A

出版信息

Med Tekh. 1997 Sep-Oct(5):11-4.

PMID:9445971
Abstract

Depth doses for radiation of superficial x-ray therapy unit RTA were measured in tissue-equivalent medium using an ionizing chamber. The measurements were made for high voltage (30-80 kV), treatment cone 8 cm in diameter and source-surface distance 30 cm. Necessary correction of the results was performed and the estimated error on a half-dose level was equal to 5.3% in 95% confidence interval. The results were in good agreement with Markus and Kasten's data but in disagreement with the data from current guide of practice on superficial x-ray therapy and the difference in half-dose level was up to two times.

摘要

使用电离室在组织等效介质中测量了浅表X射线治疗装置RTA的深度剂量。测量针对高电压(30 - 80 kV)、直径8 cm的治疗锥体以及源皮距30 cm进行。对结果进行了必要的校正,在95%置信区间内,半剂量水平的估计误差等于5.3%。结果与马库斯和卡斯滕的数据高度一致,但与当前浅表X射线治疗实践指南中的数据不一致,半剂量水平的差异高达两倍。