Suppr超能文献

使用直接数字化牙科X射线系统检测质量上的微小差异。

Detection of small differences in mass using a direct digital dental X-ray system.

作者信息

Chen S K, Hollender L, Omnell K A

机构信息

Division of Oral Radiology, School of Dentistry, National Taiwan University, Taipei.

出版信息

Dentomaxillofac Radiol. 1997 Jan;26(1):63-6. doi: 10.1038/sj.dmfr.4600202.

Abstract

OBJECTIVES

To evaluate experimentally how in direct digital radiography heterogenous detector response may influence the depiction of small mass differences.

METHODS

A step wedge of a bone-simulating resin was imaged at 70 kVp using RadioVisioGraphy ZHR 32000. Exposure time and step wedge position on the detector varied. The pixel values of the stored images of the step wedge were converted to luminance values of the displayed images.

RESULTS

The relationship between pixel and luminance values was non-linear. The difference in luminance between adjacent steps varied with position of step wedge. Regardless of position, there was an increase in difference of luminance between steps with increasing exposure. A greater relative change in luminance was consequently seen in the thinner parts of the step wedge. The failure to perceive individual steps of the wedge in its thicker parts was related to insufficient relative change in luminance between steps.

CONCLUSIONS

The heterogeneous response of the detector resulted in a varying ability to depict small mass changes over the detector area.

摘要

目的

通过实验评估在直接数字X线摄影中,探测器的异质响应如何影响小质量差异的显示。

方法

使用RadioVisioGraphy ZHR 32000在70 kVp下对骨模拟树脂阶梯楔形块进行成像。探测器上的曝光时间和阶梯楔形块位置有所变化。将阶梯楔形块存储图像的像素值转换为显示图像的亮度值。

结果

像素值与亮度值之间的关系是非线性的。相邻阶梯之间的亮度差异随阶梯楔形块的位置而变化。无论位置如何,随着曝光增加,阶梯之间的亮度差异都会增大。因此,在阶梯楔形块较薄的部分可以看到更大的亮度相对变化。在阶梯楔形块较厚的部分无法分辨各个阶梯,这与阶梯之间亮度的相对变化不足有关。

结论

探测器的异质响应导致在探测器区域描绘小质量变化的能力各不相同。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验