Samei E, Flynn M J, Reimann D A
Department of Diagnostic Radiology, Henry Ford Health System, Detroit, Michigan 48202, USA.
Med Phys. 1998 Jan;25(1):102-13. doi: 10.1118/1.598165.
The modulation transfer function (MTF) of radiographic systems is frequently evaluated by measuring the system's line spread function (LSF) using narrow slits. The slit method requires precise fabrication and alignment of a slit and high radiation exposure. An alternative method for determining the MTF uses a sharp, attenuating edge device. We have constructed an edge device from a 250-microm-thick lead foil laminated between two thin slabs of acrylic. The device is placed near the detector and aligned with the aid of a laser beam and a holder such that a polished edge is parallel to the x-ray beam. A digital image of the edge is processed to obtain the presampled MTF. The image processing includes automated determination of the edge angle, reprojection, sub-binning, smoothing of the edge spread function (ESF), and spectral estimation. This edge method has been compared to the slit method using measurements on standard and high-resolution imaging plates of a digital storage phosphor (DSP) radiography system. The experimental results for both methods agree with a mean MTF difference of 0.008. The edge method provides a convenient measurement of the presampled MTF for digital radiographic systems with good response at low frequencies.
射线照相系统的调制传递函数(MTF)通常通过使用窄缝测量系统的线扩展函数(LSF)来评估。狭缝法需要精确制造和对准狭缝,并且辐射暴露量高。一种确定MTF的替代方法是使用锐利的衰减边缘装置。我们用夹在两块薄丙烯酸板之间的250微米厚的铅箔构建了一个边缘装置。将该装置放置在探测器附近,并借助激光束和支架进行对准,使抛光边缘与X射线束平行。对边缘的数字图像进行处理以获得预采样的MTF。图像处理包括自动确定边缘角度、重投影、子分箱、边缘扩展函数(ESF)的平滑处理以及频谱估计。已使用数字存储磷光体(DSP)射线照相系统的标准和高分辨率成像板上的测量结果,将这种边缘法与狭缝法进行了比较。两种方法的实验结果相符,平均MTF差异为0.008。边缘法为数字射线照相系统的预采样MTF提供了一种方便的测量方法,在低频下具有良好的响应。