O'Brien P, Hopkins L, McCarthy D, Murphy S
Department of Haematology and Blood Transfusion, St. Vincent's Hospital, Dublin, Ireland.
Vox Sang. 1998;74(1):53-5.
An anti-Jka was not detectable by the DiaMed-ID Micro Typing System but was strongly reactive when tested using other indirect antiglobulin test (IAT) methods. However, it was detectable in the DiaMed-ID system when the recommended diluent ID-Diluent 2, was replaced by other low ionic Strength saline (LISS) solutions. This led us to believe that the problem lay with the ID-Diluent 2.
The anti-Jka was investigated using various low ionic strength diluents in the IAT against both polyspecific and monospecific antiglobulin reagents. We tested the ID-Diluent 2 for its ability to inhibit complement activation, and bind calcium.
The anti-Jka was shown to be a 'complement-only' binding antibody that reacted in all IATs performed except when tested with ID-Diluent 2. The ID-Diluent was shown to prevent haemolysis of ABO lytic antibodies, and further testing shows that it binds calcium.
The use of a diluent which prevents haemolysis and binds calcium in an IAT system that contains anti-complement as well as anti-IgG must be questioned.
DiaMed-ID微柱凝集系统检测不到抗Jka,但使用其他间接抗球蛋白试验(IAT)方法检测时呈强反应性。然而,当推荐的稀释液ID-Diluent 2被其他低离子强度盐水(LISS)溶液取代时,在DiaMed-ID系统中可检测到抗Jka。这使我们认为问题出在ID-Diluent 2上。
在IAT中使用各种低离子强度稀释液针对多特异性和单特异性抗球蛋白试剂对抗Jka进行研究。我们测试了ID-Diluent 2抑制补体激活和结合钙的能力。
抗Jka被证明是一种“仅补体”结合抗体,在所有进行的IAT中均有反应,但用ID-Diluent 2检测时除外。ID-Diluent被证明可防止ABO溶血性抗体的溶血,进一步检测表明它能结合钙。
在含有抗补体以及抗IgG的IAT系统中,使用能防止溶血并结合钙的稀释液这一做法必须受到质疑。