Ultrasonic blood flow measurements: transcutaneous compared with implanted cuff.

作者信息

Greene E R, Histand M B, Miller C W

出版信息

ISA Trans. 1976;15(1):88-94.

PMID:950285
Abstract

Experiments were undertaken in our laboratory to establish relationships between flow parameters calculated from implanted ultrasonic cuff signals and those obtained with a transcutaneous probe. Analog signals were received from crystals surgically implanted on the right carotid, right femoral, and abdominal aorta of a mature male beagle dog. Near these locations, transcutaneous measurements were made using a prescribed technique. From these ultrasonic signals, the following time dependent flow parameters were calculated: spatial velocities, velocity profiles, diameters, flows, and velocity frequency distributions. Using these measurements, a comparison between the implanted cuff and transcutaneous probe was made. In addition to high reproductibility, results confirm a strong correlation between the two techniques; consequently, the validity of the less traumatic transcutaneous method is substantiated.

摘要

我们实验室进行了多项实验,以确定根据植入式超声袖带信号计算出的血流参数与经皮探头所获血流参数之间的关系。从手术植入一只成年雄性比格犬的右颈动脉、右股动脉和腹主动脉的晶体接收模拟信号。在这些位置附近,采用规定技术进行经皮测量。从这些超声信号中,计算出以下随时间变化的血流参数:空间速度、速度分布、直径、流量和速度频率分布。利用这些测量结果,对植入式袖带和经皮探头进行了比较。除了具有高度可重复性外,结果证实了这两种技术之间存在很强的相关性;因此,创伤较小的经皮方法的有效性得到了证实。