Re-used pacemakers--as safe as new? A retrospective case-control study.

作者信息

Linde C L, Bocray A, Jonsson H, Rosenqvist M, Rådegran K, Rydén L

机构信息

Department of Cardiology, Karolinska Hospital, Stockholm, Sweden.

出版信息

Eur Heart J. 1998 Jan;19(1):154-7. doi: 10.1053/euhj.1997.0728.

Abstract

UNLABELLED

Re-use of pacemakers is of interest in an era of declining health care resources if it is proven safe and without risks to the patients. In order to investigate the safety of re-use of pacemakers we performed a retrospective case-control study. One hundred patients, who received a re-used pacemaker, were matched for date of implantation and mode (AAI; VVI; DDD) to 100 others who received a newly manufactured pacemaker and were followed for a mean of 32 +/- 11 months for complications defined as infections and signs of pacemaker malfunction and pacemaker replacement earlier than anticipated due to battery depletion. Patients who received re-used pacemakers were significantly older than those who received new units (79 +/- 9 vs 68 +/- 21 years; P < 0.0001). The number of complications did not differ significantly between groups. There were no early replacements due to battery depletion in either of the two groups. A cost-benefit analysis revealed a substantial economical advantage.

CONCLUSIONS

The re-use of pacemakers can be carried out without increased risk to the patients provided a proper routine for technical control and sterilization is followed. Re-use means substantial savings which possibly could make advanced pacemaker treatment available to all eligible patients irrespective of age. Whether re-use is feasible with implantable defibrillators remains to be determined.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验