O'Bannon W, Donachy J H, Brighton J A, Phillips W M, Prophet G A, Ferrari O, Pierce W S
Trans Am Soc Artif Intern Organs. 1976;22:450-9.
Long-term synchronized left ventricular bypass has been performed in calves using pneumatically powered pumps having a smooth lining fabricated of segmented polyurethane. Three different pump designs have been employed: a) sac pump, b) longitudinal tethered sac pump, and c) transverse tethered sac pump. Sizeable thrombi occurred in the apex of the sac and longitudinal tethered sac pumps. In the transverse tethered sac pumps, a considerably smaller thrombus was seen only at the flexion point on the inlet side of the pump. Thrombi occurred in areas of low flow velocity and were not prevented by anticoagulation or surface coating techniques. This study indicates the importance of ventricle design in the development of thrombus free pumps and suggests modifications to the present pumps to reduce or eliminate thrombus formation.
使用由分段聚氨酯制成的具有光滑内衬的气动泵,在小牛身上进行了长期同步左心室旁路手术。采用了三种不同的泵设计:a)囊式泵,b)纵向系留囊式泵,c)横向系留囊式泵。囊式泵和纵向系留囊式泵的顶端出现了大量血栓。在横向系留囊式泵中,仅在泵入口侧的弯曲点处发现了一个小得多的血栓。血栓出现在低流速区域,抗凝或表面涂层技术无法预防。这项研究表明心室设计在无血栓泵的开发中的重要性,并建议对现有泵进行改进,以减少或消除血栓形成。