气袋导致起搏器故障。

Gas pocket causing pacemaker malfunction.

作者信息

Lasala A F, Fieldman A, Diana D J, Humphrey C B

出版信息

Pacing Clin Electrophysiol. 1979 Mar;2(2):183-5. doi: 10.1111/j.1540-8159.1979.tb05198.x.

Abstract

Pacemaker malfunction was attributed to the increase in impedance to current flow caused by a pocket of air separating the anodal contact plate of a unipolar generator from the overlying skin. Lack of capture was noted 20 hours after implantation. The malfunction was permanently corrected by bedside aspiration of the gas with a sterile syringe.

摘要

起搏器故障归因于单极发生器的阳极接触板与覆盖皮肤之间的气袋导致电流流动阻抗增加。植入后20小时发现无法捕获。通过床边用无菌注射器抽吸气体,故障得到了永久性纠正。