MacDougal C S, Rigas M L, Ben-Jebria A, Ultman J S
Department of Chemical Engineering, Pennsylvania State University, University Park 16802, USA.
Arch Environ Health. 1998 Mar-Apr;53(2):161-74. doi: 10.1080/00039896.1998.10545978.
The breath-to-breath determination of total respiratory ozone (O3) uptake requires the monitoring of O3 concentration at the airway opening with an instrument that responds rapidly relative to the frequency of respiration. Originally, the authors developed an analyzer that used the homogeneous chemiluminescent reaction of O3 with 2-methyl-2-butene, but it was suitable only for monitoring O3 during quiet breathing and light exercise (Ben-Jebria and Ultman, Rev Sci Instrum 1989; 60:3004-11, and Ben-Jebria et al., Rev Sci Instrum 1990; 61:3435-39). The improvement of performance characteristics of the aforementioned analyzer enabled the authors to use the newly constructed and self-contained instrument, which used ethylene as the reactant gas, for respiratory O3 monitoring during moderate-to-heavy exercise. Operating at a reaction chamber pressure of 350 torr, an ethylene/sample flow ratio of 4:1, and a sampling flow of 0.6 lpm, the authors achieved an optimum analyzer performance (i.e., 10-90% step-response of 70 msec and a minimum resolution of 0.006 ppm O3). Furthermore, the new instrument did not exhibit the nonlinear calibration and the CO2 interference suffered by the original analyzer. To demonstrate the quality of the new O3 analyzer in a respiratory application (i.e., total O3 uptake), the authors measured a series of single breaths on two subjects who breathed 0.11 and 0.43 ppm O3-in-air mixtures for 15 min during rest, and during moderate and heavy exercise.
逐次呼吸测定总呼吸性臭氧(O₃)摄取量需要使用一种相对于呼吸频率响应迅速的仪器来监测气道开口处的O₃浓度。最初,作者开发了一种分析仪,它利用O₃与2-甲基-2-丁烯的均相化学发光反应,但它仅适用于安静呼吸和轻度运动时的O₃监测(Ben-Jebria和Ultman,《科学仪器评论》1989年;60:3004 - 11,以及Ben-Jebria等人,《科学仪器评论》1990年;61:3435 - 39)。上述分析仪性能特征的改进使作者能够使用新构建的独立仪器,该仪器以乙烯作为反应气体,用于中度至重度运动时的呼吸O₃监测。在反应室压力为350托、乙烯/样品流量比为4:1以及采样流量为0.6升/分钟的条件下运行,作者实现了最佳分析仪性能(即10 - 90%阶跃响应为70毫秒,O₃的最小分辨率为0.006 ppm)。此外,新仪器没有出现原始分析仪所遭受的非线性校准和CO₂干扰问题。为了证明新型O₃分析仪在呼吸应用(即总O₃摄取量)中的质量,作者对两名受试者进行了一系列单次呼吸测量,这两名受试者在休息、中度和重度运动期间吸入含0.11和0.43 ppm O₃的空气混合物15分钟。