Wheeler D L, McLoughlin S W
Department of Orthopaedics, University of Florida, Gainesville 32610, USA.
Clin Orthop Relat Res. 1998 May(350):237-45.
A series of mechanical tests on anatomic specimen cancellous bone and cancellous bonelike foam were conducted to evaluate and compare an Acutrak compression screw with an AO 4-mm cancellous screw and the Herbert screw. The Acutrak and AO screws produced similar fragment compression in foam and bone; Acutrak and AO compression were significantly greater than that of the Herbert screw. However, Acutrak was able to maintain compression after cyclic loading significantly better than were the AO and Herbert screws. The pushout force of Acutrak and AO screws were significantly greater than that of the Herbert screw in foam and bone. The torque required to break fragment contact was significantly greater for the Acutrak than for the AO or Herbert screws, but that for the AO was greater than that for the Herbert screw. The results of this analysis show the Acutrak screw is capable of producing and maintaining compression between bone fracture fragments. In addition, the Acutrak screw was shown to have superior mechanical characteristics than did Herbert screw in every mode tested. The Acutrak screw did not surpass the fragment compression achieved by the AO screw in foam, but it did not overtighten or experience degradation of compression after 500 cycles of simulated physiologic loading.
对解剖标本松质骨和类松质骨泡沫材料进行了一系列力学测试,以评估和比较Acutrak加压螺钉与AO 4毫米松质骨螺钉及Herbert螺钉。Acutrak螺钉和AO螺钉在泡沫材料和骨中产生的骨折块压缩效果相似;Acutrak螺钉和AO螺钉产生的压缩力明显大于Herbert螺钉。然而,在循环加载后,Acutrak螺钉保持压缩的能力明显优于AO螺钉和Herbert螺钉。在泡沫材料和骨中,Acutrak螺钉和AO螺钉的推出力明显大于Herbert螺钉。使骨折块接触分离所需的扭矩,Acutrak螺钉明显大于AO螺钉或Herbert螺钉,但AO螺钉大于Herbert螺钉。该分析结果表明,Acutrak螺钉能够在骨折块之间产生并维持压缩力。此外,在每种测试模式下,Acutrak螺钉均显示出比Herbert螺钉更优异的力学特性。在泡沫材料中,Acutrak螺钉产生的骨折块压缩力未超过AO螺钉,但在500次模拟生理负荷循环后,它不会过度拧紧或出现压缩力下降的情况。