Suppr超能文献

Yasargil aneurysm clips: evaluation of interactions with a 1.5-T MR system.

作者信息

Shellock F G, Kanal E

机构信息

Department of Radiology, University of Southern California, School of Medicine, Los Angeles, USA.

出版信息

Radiology. 1998 Jun;207(3):587-91. doi: 10.1148/radiology.207.3.9609877.

Abstract

PURPOSE

To evaluate the magnetic field interactions with Yasargil aneurysm clips exposed to a 1.5-T MR system.

MATERIALS AND METHODS

Sixteen Yasargil aneurysm clips made with Phynox (10 clips returned to the manufacturer due to suspected magnetic field interactions and six clips obtained from the manufacturer in sterile packages) were tested for magnetic field interaction at 1.5 T by using techniques to assess magnetic field-induced translational (deflection angle test) and torque (changes in alignment or rotation) forces.

RESULTS

None of the aneurysm clips demonstrated magnetic field interactions; that is, deflection angles were zero and there were no changes in alignment or evidence of rotational forces present.

CONCLUSION

It is safe to perform MR procedures at 1.5 T or less in patients with Yasargil aneurysm clips made with nonferromagnetic material.