克里斯蒂安森内置假体的一种新设计。两例假体头部两个组件分离病例的报告。

A new design of the Christiansen endoprosthesis. Report of two cases in which separation of the two components of the prosthetic head occurred.

作者信息

Solhaug J H, Molster A, Soreide O

出版信息

Acta Orthop Scand. 1976 Aug;47(4):414-8.

PMID:961397
Abstract

Primary prosthetic replacement in elderly patients with displaced intracapsular fracture of the femoral neck has been used during the last few years with encouraging early results. The prosthetic device preferred has been Christiansen's endoprosthesis. During the past year a new type with separable plastic head and metallic cap has been used. This modification has resulted in the occurrence of a previously unknown complication, a separation of these two parts, necessitating reoperation and insertion of a new head piece. The possible mechanisms causing this separation are discussed.