持续动静脉血液透析滤过:AN - 69 型毛细血管血液滤过器的水力和扩散通透指数

Continuous arterio-venous haemodiafiltration: hydraulic and diffusive permeability index of an AN-69 capillary haemofilter.

作者信息

Akcahuseyin E, van Duyl W A, Vincent H H, Vos M C, Schalekamp M A

机构信息

Department of Internal Medicine I, Faculty of Medicine, Erasmus University, Rotterdam, The Netherlands.

出版信息

Med Biol Eng Comput. 1998 Jan;36(1):43-50. doi: 10.1007/BF02522856.

Abstract

The dependence of uraemic solute clearance on the hydraulic and diffusive permeability index of an AN-69 capillary haemofilter is investigated during the treatment of patients with continuous arterio-venous haemodiafiltration (CAVHD). A mathematical model is presented to calculate solute clearance and the hydraulic and diffusive permeability index parameters from clinical data and to predict the blood flow rate entering the extra-corporeal circuit from the manufacturer's specifications and blood viscosity. By measuring the flow rates, the patient's mean arterio-venous pressure difference and uraemic solute clearance under different clinical and operational conditions, mathematical model equations are evaluated. During the average survival time of an AN-69 capillary haemofilter of about five days, it is found that both the hydraulic permeability index and the diffusive permeability index decline over treatment time, independent of the haemofilter resistance to blood flow. The measured haemofilter resistance to blood flow is three times higher than the haemofilter resistance predicted from the manufacturer's specifications and blood viscosity. Predicting the blood flow rate entering the extra-corporeal circuit from the arterial haematocrit, plasma protein concentration and temperature and the manufacturer's specifications is not reliable.

摘要

在对连续动静脉血液透析滤过(CAVHD)患者进行治疗期间,研究了尿毒症溶质清除率对AN - 69毛细管血液滤过器水力学和扩散渗透指数的依赖性。提出了一个数学模型,用于根据临床数据计算溶质清除率以及水力学和扩散渗透指数参数,并根据制造商的规格和血液粘度预测进入体外循环的血流速度。通过测量不同临床和操作条件下的流速、患者平均动静脉压差和尿毒症溶质清除率,对数学模型方程进行了评估。在AN - 69毛细管血液滤过器约五天的平均使用寿命期间,发现水力学渗透指数和扩散渗透指数均随治疗时间下降,且与血液滤过器对血流的阻力无关。测得的血液滤过器对血流的阻力比根据制造商的规格和血液粘度预测的血液滤过器阻力高三倍。根据动脉血细胞比容、血浆蛋白浓度和温度以及制造商的规格预测进入体外循环的血流速度并不可靠。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验