Pastor S J, Wood S H, Wilkins C L
Department of Chemistry, University of California-Riverside 92521, USA.
J Mass Spectrom. 1998 May;33(5):473-9. doi: 10.1002/(SICI)1096-9888(199805)33:5<473::AID-JMS659>3.0.CO;2-O.
Detection limits of poly(ethylene glycol) were examined in the mass range 2000-6000 Da. Using an aerospray sample deposition technique, highly uniform sample surfaces were produced. This method allows signal averaging of spectra from up to 400 shots on the same sample spot. It is found that, as the material available for desorption is decreased, the overall average sample consumption per shot is decreased. Experimentally determined detection limits of 40 and 280 fmol (based on the average molecular masses of 2000 and 6000) were found for PEG 2000 and PEG 6000, respectively. The sample spectra show oligomer distributions in agreement with their higher concentration counterparts. However, at the lowest signal-to-noise levels, oligomers at the extremes of the distribution are no longer detected, making the polymer distribution appear to be narrower in mass range.
在2000 - 6000道尔顿的质量范围内检测了聚乙二醇的检测限。采用气溶胶喷雾样品沉积技术,制备了高度均匀的样品表面。该方法允许对同一样品点上多达400次的光谱进行信号平均。结果发现,随着可用于解吸的材料减少,每次进样的总体平均样品消耗量也会降低。实验测定的聚乙二醇2000和聚乙二醇6000的检测限分别为40和280飞摩尔(基于平均分子量2000和6000)。样品光谱显示的低聚物分布与其较高浓度的对应物一致。然而,在最低信噪比水平下,分布两端的低聚物不再被检测到,使得聚合物分布在质量范围内看起来变窄。