Inappropriate sensing of atrial stimuli in patients with third-generation defibrillators and DDD pacemakers.

作者信息

Lampert R, McPherson C A, Lewis R J, Batsford W P, Rosenfeld L E

机构信息

Yale University School of Medicine, Yale-New Haven Hospital, Connecticut 06511, USA.

出版信息

Pacing Clin Electrophysiol. 1998 Jun;21(6):1225-9. doi: 10.1111/j.1540-8159.1998.tb00181.x.

Abstract

Although the problem of ICD sensing of paced ventricular stimuli has been resolved by incorporation of VVI pacing into current ICDs, many patients required separate DDD pacemakers. We report a problematic PM-ICD interaction: the inability to prevent sensing of paced atrial stimuli ("atrial sensing") leading to double-counting in DDD-PM-requiring patients with transvenous (TV) ICDs with aggressive autogain sensing (CPI Ventak PRxII or III). Four of eight patients receiving both transvenous DDD PMs and ICDs (CPI Endotak lead, at the RV apex), had atrial sensing, leading to double counting, despite intraoperative testing of multiple atrial locations with an active fixation lead. Five patients had a PRxII/III ICD, four with atrial sensing (80%), and three a PRx without atrial sensing. Patients with atrial sensing were not distinguished by any clinical or device related variable. In patients with atrial sensing (all with heart block), the PM was programmed to VDD mode. No patient has received inappropriate therapy or failed to sense VF in follow-up. In many patients with TV ICDs who required DDD pacing, no atrial position can be found without ICD sensing of atrial stimuli. While in patients with heart block this problem can be circumvented by programming to the VDD mode, in patients with sinus incompetence it may only be resolved by the combination ICD-DDD PM, currently in development.

摘要

相似文献

1
Inappropriate sensing of atrial stimuli in patients with third-generation defibrillators and DDD pacemakers.
Pacing Clin Electrophysiol. 1998 Jun;21(6):1225-9. doi: 10.1111/j.1540-8159.1998.tb00181.x.
10
Functional atrial undersensing associated with switching to a tracking mode of pacing.
Pacing Clin Electrophysiol. 2012 Oct;35(10):1188-93. doi: 10.1111/j.1540-8159.2012.03461.x. Epub 2012 Jul 21.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验