[使用圣犹达医疗公司19A-HP型人工心脏瓣膜的经验,该瓣膜在术后早期导致开口受限]

[An experience with St. Jude medical prosthetic 19A-HP which resulted in restricted opening at an early stage after operation].

作者信息

Mashiko K, Ishii S, Naganuma H, Sakamoto H, Yagi H, Seo A, Mikawa H

机构信息

Department of Cardiovascular Surgery, Fuji City Hospital, Shizuoka, Japan.

出版信息

Kyobu Geka. 1998 Jun;51(6):501-3.

PMID:9637846
Abstract

St. Jude medical hemodynamic plus series is positively used for aortic valve replacement (AVR) of small orifice because it can increase the orifice area by 26% compared with conventional prosthetic valves of the same size. We performed AVR with SJM 19A-HP on a patient having aortic stenosis with regurgitation using horizontal mattress suturing technique. The aortic orifice size was 18.9 mm at the preoperative measurement. The course after the operation was uneventful until 9 days after the operation when cinefluoroscopy revealed symmetrical restriction of opening of valve leaflets. We are now monitoring the clinical course at the patient has not developed any symptoms though the LV-Ao pressures gradient is 60 mmHg by Doppler echocardiography. As the restriction of valve leaflet opening was symmetrical, it is not likely that an excess of the ligature or remnant caused the restriction; structural problems of St. Jude medical prosthesis appear to be a more reasonable explanation. Since the orifice ring of this prosthesis is not reinforced, longitudinal forces applied to the hinge may alter the shape of the ring, thus restricting the movement of valve leaflets. In the present case, forced insertion of 19A-HP instead of more desirable 17-HP is considered to have caused longitudinal forces acting on the hinge.

摘要

圣犹达医疗血流动力学增强系列产品被积极用于小瓣口的主动脉瓣置换术(AVR),因为与相同尺寸的传统人工瓣膜相比,它可使瓣口面积增加26%。我们使用水平褥式缝合技术,对一名患有主动脉瓣狭窄并反流的患者植入圣犹达医疗19A-HP进行主动脉瓣置换术。术前测量主动脉瓣口大小为18.9毫米。术后过程平稳,直到术后9天,荧光透视显示瓣叶开口出现对称性受限。尽管通过多普勒超声心动图测量左心室-主动脉压力梯度为60毫米汞柱,但患者目前未出现任何症状,我们正在对其临床过程进行监测。由于瓣叶开口受限是对称的,不太可能是结扎过度或残留导致的;圣犹达医疗假体的结构问题似乎是更合理的解释。由于该假体的瓣口环未加固,作用于铰链的纵向力可能会改变环的形状,从而限制瓣叶的运动。在本病例中,强行植入19A-HP而非更合适的17-HP被认为是导致作用于铰链的纵向力的原因。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验