Khaksa G, Nalini K, Bhat M, Udupa N
Department of Pharmaceutics, College of Pharmaceutical Sciences, Manipal, Karnataka State, India.
Anal Biochem. 1998 Jun 15;260(1):92-5. doi: 10.1006/abio.1998.2660.
A rapid and simple isocratic chromatographic procedure for the determination of insulin in human and rat plasma using reversed-phase (RP) high-performance liquid chromatography with an ultraviolet/visible detector is described. The method includes extraction of insulin from human and rat plasma into dichloromethane, followed by back-extraction into 0.05 N hydrochloric acid. The organic phase was evaporated under a stream of nitrogen. The aqueous phase was filtered and a 100 microliters aliquot was analyzed on a RP-C18 column eluted with a mobile phase consisting of a mixture of 74 vol of 0.2 M sodium sulfate anhydrous adjusted to pH 2.3 with phosphoric acid and 26 vol of acetonitrile. The flow rate was 1.2 ml/min and the wavelength was set at 214 nm. The calibration curve was linear over the range of 75-800 microIU/ml. The precision of the assay expressed as coefficients of variation was less than 6% over the entire concentration range. The recovery for insulin ranged from 79 to 81% from human and rat plasma, with coefficients of variation less than 6%. The intra- and interassay coefficients of variation were less than 5.7%. The limit of detection was 50 microIU/ml.
描述了一种使用反相(RP)高效液相色谱法和紫外/可见检测器测定人和大鼠血浆中胰岛素的快速简单等度色谱方法。该方法包括将人和大鼠血浆中的胰岛素萃取到二氯甲烷中,然后反萃取到0.05N盐酸中。有机相在氮气流下蒸发。水相过滤后,取100微升等分试样在RP-C18柱上进行分析,流动相由74体积的用磷酸调节至pH 2.3的0.2M无水硫酸钠和26体积的乙腈混合而成。流速为1.2ml/min,波长设定为214nm。校准曲线在75-800微国际单位/毫升范围内呈线性。在整个浓度范围内,以变异系数表示的测定精密度小于6%。人和大鼠血浆中胰岛素的回收率在79%至81%之间,变异系数小于6%。批内和批间变异系数小于5.7%。检测限为50微国际单位/毫升。