Variations in depth-dose data between open and wedge fields for 4-MV X-rays.

作者信息

Sewchand W, Khan F M, Williamson J

出版信息

Radiology. 1978 Jun;127(3):789-92. doi: 10.1148/127.3.789.

Abstract

Central-axis depth-dose data for 4-MV x rays, including tissue-maximum ratios, were measured for wedge fields. Comparison with corresponding open-field data revealed differences in magnitude which increased with depth, field size, and wedge thickness. However, phantom scatter correction factors for the wedge fields differed less than 1% from corresponding open-field factors. The differences in central-axis per cent depth doses between the two types of fields indicate beam hardening by the wedge filter. This study also implies that the derivation of tissue-maximum ratios from central-axis per cent depth is as valid for wedge as for open fields.