Wada H, Koike T, Kobayashi T
Department of Mechanical Engineering, Tohoku University, Sendai, Japan.
Ear Hear. 1998 Jun;19(3):240-9. doi: 10.1097/00003446-199806000-00007.
To develop a measuring apparatus that has the ability to sweep both frequency and external auditory meatus static pressure and can display measurement results in a three-dimensional expression. To measure the middle ear dynamic characteristics of normal-hearing subjects and of patients with this apparatus.
Investigate 275 ears of 153 normal-hearing subjects and 72 ears with middle ear diseases.
The measurement results show fairly distinctive patterns depending on the middle ear conditions, i.e., normal, ossicular chain separation, ossicular chain fixation, secretory otitis media, tympanic membrane perforation, and tympanic membrane atelectasis. The evaluation of patients with ossicular diseases revealed that the rate of correct diagnosis of ossicular chain separation is 84% and that of ossicular chain fixation is 74%. These diagnoses were confirmed by surgery.
Displaying the measurement results in a three-dimensional expression is helpful to make correct diagnosis in clinical practice. It is especially easy to make a distinction between the ossicular chain separation and ossicular chain fixation. Therefore, it is concluded that this apparatus has a high degree of clinical applicability to the diagnosis of these diseases.
研发一种能够同时扫描频率和外耳道静压,并能以三维形式显示测量结果的测量装置。使用该装置测量听力正常受试者及患者的中耳动态特性。
对153名听力正常受试者的275只耳朵以及72只患有中耳疾病的耳朵进行研究。
测量结果根据中耳状况呈现出相当独特的模式,即正常、听骨链分离、听骨链固定、分泌性中耳炎、鼓膜穿孔和鼓膜萎缩。对听骨疾病患者的评估显示,听骨链分离的正确诊断率为84%,听骨链固定的正确诊断率为74%。这些诊断均通过手术得到证实。
以三维形式显示测量结果有助于在临床实践中做出正确诊断。尤其容易区分听骨链分离和听骨链固定。因此,得出结论,该装置对这些疾病的诊断具有高度的临床适用性。