Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children.

作者信息

Kalifa G, Charpak Y, Maccia C, Fery-Lemonnier E, Bloch J, Boussard J M, Attal M, Dubousset J, Adamsbaum C

机构信息

Service de Radiologie, Hôpital Saint-Vincent de Paul, F-75 674 Paris Cedex 14, France.

出版信息

Pediatr Radiol. 1998 Jul;28(7):557-61. doi: 10.1007/s002470050413.

Abstract

BACKGROUND

A new low-dose digital X-ray device, based on Charpak's Nobel prize-winning multiwire chamber, enables the production of images at very low doses. Objectives. To present the first dosimetric and clinical results.

MATERIALS AND METHODS

The analysis was performed on 93 children with scoliosis and 47 undergoing pelvic radiography. The comparative study between conventional X-ray and the new technique focused on three points: (1) the dose delivered by each system (2) the diagnostic information provided by each system and (3) comparison of image quality criteria with European guidelines.

RESULTS

The mean ratio of conventional dose to that of the low-dose technique was 13.1 for the spinal examination and 18.8 for the pelvis. There was no significant difference in diagnostic information available from each modality, but there was a slight difference in quality criteria in favour of the conventional technique.

CONCLUSION

This new device allows spectacular dose reduction, consistent with adequate clinical information. Improvements of the prototype will lead to extension of potential indications and industrial development.

摘要

背景

一种基于夏尔帕克获诺贝尔奖的多丝正比室的新型低剂量数字X射线设备,能够以极低剂量生成图像。目的:展示首批剂量测定和临床结果。

材料与方法

对93例脊柱侧弯儿童和47例接受骨盆X线摄影的患者进行了分析。传统X射线与新技术之间的对比研究集中在三点:(1)每个系统的剂量;(2)每个系统提供的诊断信息;(3)将图像质量标准与欧洲指南进行比较。

结果

脊柱检查中传统剂量与低剂量技术剂量的平均比值为13.1,骨盆检查为18.8。各检查方式提供的诊断信息无显著差异,但在质量标准上存在细微差异,传统技术更具优势。

结论

这种新设备可大幅降低剂量,同时提供足够的临床信息。对该原型的改进将扩大潜在适应症范围并推动产业发展。