Willson V J, Preston J R, Lockley W J, Singh J, Wilkinson D
Department of Physical and Metabolic Sciences, Astra Charnwood, Loughborough, UK.
J Pharm Biomed Anal. 1998 Aug;17(4-5):793-800. doi: 10.1016/s0731-7085(97)00272-0.
A radioimmunoassay has been developed for the determination of AR-C15849KF, a CCK-8 analogue, in human plasma. The method incorporates solid-phase sample extraction, is suitable for the determination of the analyte at pg ml-1 concentrations and is based on a method developed and validated for dog plasma. The solid-phase extraction, using ion-exchange aminopropyl and octadecyl sorbents sequentially, was retained for this procedure to remove matrix interferences in the plasma and to enhance method sensitivity. The calibration range is 10-500 pg ml-1, using a 1 ml sample of undiluted human plasma. The method has been successfully used to generate early human pharmacokinetic data during a programme of exploratory development.
已开发出一种放射免疫分析法,用于测定人血浆中的CCK-8类似物AR-C15849KF。该方法采用固相样品萃取,适用于测定皮克/毫升浓度的分析物,且基于一种已针对犬血浆开发并验证的方法。该程序保留了依次使用离子交换氨丙基和十八烷基吸附剂的固相萃取法,以去除血浆中的基质干扰并提高方法灵敏度。使用1毫升未稀释的人血浆样品,校准范围为10 - 500皮克/毫升。在探索性开发计划中,该方法已成功用于生成早期人体药代动力学数据。