Stoev G
Central Laboratory for Veterinary Control and Ecology, Sofia, Bulgaria.
J Chromatogr B Biomed Sci Appl. 1998 Jun 12;710(1-2):234-8. doi: 10.1016/s0378-4347(98)00114-5.
The influence of the pH of the mobile phase with some modifiers on the chromatographic behavior and fluorescence properties of closantel have been investigated. At acidic pH values (2-6), the benzamide moiety of the closantel forms a six-membered ring by hydrogen bonding and possesses a native fluorescence. Using the fluorescence emission of closantel at lambda(ex) = 335 nm, lambda(em) = 510 nm, and pH 2.5 of the mobile phase, a linear calibration curve was estimated over a concentration range of about two orders of magnitude with a correlation coefficient larger than 0.992. The limit of the fluorescence detection was 10 microg/kg. This value was at least 10 times lower than that using UV detection. The method was applied to the determination of closantel in plasma and tissue samples, purified by a solid-phase extraction with C18 cartridges.
研究了流动相的pH值及某些改性剂对氯氰碘柳胺色谱行为和荧光性质的影响。在酸性pH值(2 - 6)下,氯氰碘柳胺的苯甲酰胺部分通过氢键形成六元环并具有天然荧光。利用氯氰碘柳胺在激发波长λ(ex)=335nm、发射波长λ(em)=510nm以及流动相pH值为2.5时的荧光发射,在约两个数量级的浓度范围内估算出线性校准曲线,相关系数大于0.992。荧光检测限为10μg/kg。该值至少比使用紫外检测时低10倍。该方法应用于通过C18柱固相萃取纯化后的血浆和组织样品中氯氰碘柳胺的测定。