Kasuya Y, Kawai H, Yamamoto T, Dohi S
Department of Anesthesiology and Critical Care Medicine, Gifu University School of Medicine.
Masui. 1998 Jun;47(6):756-8.
We measured cardiac outputs at forty points from five patients by pulse dye densitometry and compared these with those measured by thermodilution method. We obtained a good correlation (y = 1.090 x = . 0.030, n = 40, 5 cases) and small mean bias (0.348 +/- 0.830 l.min-1, n = 40, 5 cases) between the two methods. We suggest that this method of cardiac output measurement by pulse dye densitometry is not invasive and may be more effective than that by thermodilution method.
我们通过脉冲染料密度测定法在5名患者的40个点测量了心输出量,并将其与通过热稀释法测量的结果进行比较。我们发现两种方法之间具有良好的相关性(y = 1.090x + 0.030,n = 40,5例),平均偏差较小(0.348 +/- 0.830 l.min-1,n = 40,5例)。我们认为,这种通过脉冲染料密度测定法测量心输出量的方法是非侵入性的,可能比热稀释法更有效。