Kilár F, Végvári A, Mód A
Central Research Laboratory, Medical School, University of Pécs, Hungary.
J Chromatogr A. 1998 Jul 17;813(2):349-60. doi: 10.1016/s0021-9673(98)00377-x.
Substituted aminomethylphenol dyes, low-molecular-mass isoelectric point (pI) markers and hemoglobin samples from normal individuals and diabetic patients were used to test a new set-up of capillary isoelectric focusing (cIEF) in uncoated capillaries. In previous cIEF methods, a mixture of sample components and carrier ampholytes was applied in the capillary and analyzed. In the new set-up a fractionated injection protocol is used to apply a 'sandwich' ampholyte-sample-ampholyte plug in the capillary for analysis. This new set-up allows the separation of amphoteric compounds having pI values outside the pH region of the ampholytes applied in the capillary with high precision. The high resolution power of this technique was proven with the analysis of hemoglobin variants.
使用取代氨甲基苯酚染料、低分子量等电点(pI)标记物以及来自正常个体和糖尿病患者的血红蛋白样本,在未涂层毛细管中测试了一种新型毛细管等电聚焦(cIEF)装置。在以往的cIEF方法中,将样品成分和载体两性电解质的混合物注入毛细管进行分析。在新装置中,采用分步进样方案,在毛细管中注入一个“三明治”式的两性电解质-样品-两性电解质塞进行分析。这种新装置能够高精度地分离毛细管中所施加两性电解质pH区域之外pI值的两性化合物。通过对血红蛋白变体的分析证明了该技术的高分辨率。