Hsu E W, Reeder S B, MacFall J R
Center for In Vivo Microscopy, Duke University Medical Center, Durham, North Carolina 27710, USA.
Magn Reson Med. 1998 Aug;40(2):334-40. doi: 10.1002/mrm.1910400220.
Adiabatic pulses, although useful in generating uniform spin nutation in the presence of inhomogeneous B1 fields, are limited for NMR imaging applications due to the lack of slice-selective excitation capability. Selective excitation techniques using gradient modulation have been introduced; however, present methods require either a minimum of two excitations or eight adiabatic segments. Here, a scheme is presented that allows single-shot, arbitrary flip-angle, and slice-selective excitation with only four adiabatic half-passage segments. The technique is demonstrated via computer simulation and experimental tests on a phantom. Furthermore, issues associated with the implementation of these gradient-modulated adiabatic pulses are discussed.