Ahrenholz P, Winkler R E, Ramlow W, Tiess M, Thews O
BioArtProducts GmbH, Rostock, Germany.
Int J Artif Organs. 1998 Jun;21(6):321-7.
With on-line formation of the substitution fluid, high substitution rates in predilution (PRD) and postdilution (POD) can be obtained (Fresenius 4008 On-Line HDF; Gambro AK 100 Ultra). The substitution fluid is branched off from the dialysate produced by the dialysate delivery system of the HDF machine. Under these conditions it is desirable to consider the effect of the different treatment modes on the acid-base status. Using Fresenius 4008 On-Line HDF machines, ESRD-patients were treated alternately with high-flux hemodialysis (HD), postdilution HDF (POD-HDF) and predilution HDF (PRD-HDF), while all other treatment parameters were kept constant, in particular the bicarbonate dialysate concentration. Plasma-HCO3, -pH and -pCO2 were measured and compared with the results of a multicompartment bicarbonate model developed by Thews. Also plasma-pO2 and K+ were measured. The results showed no significant differences between HD, POD- and PRD-HDF. Acidosis was corrected effectively and no excessive compensation of the acid-base disturbance was observed.
通过在线生成置换液,可以在预稀释(PRD)和后稀释(POD)中获得高置换率(费森尤斯4008在线血液透析滤过;金宝AK 100 Ultra)。置换液从血液透析滤过机的透析液输送系统产生的透析液中分流出来。在这些条件下,需要考虑不同治疗模式对酸碱状态的影响。使用费森尤斯4008在线血液透析滤过机,对终末期肾病患者交替进行高通量血液透析(HD)、后稀释血液透析滤过(POD-HDF)和预稀释血液透析滤过(PRD-HDF)治疗,同时保持所有其他治疗参数不变,尤其是碳酸氢盐透析液浓度。测量血浆-HCO3、-pH和-pCO2,并与Thews开发的多室碳酸氢盐模型的结果进行比较。还测量了血浆-pO2和K+。结果显示HD、POD-HDF和PRD-HDF之间无显著差异。酸中毒得到有效纠正,未观察到酸碱紊乱的过度代偿。