Smit J L, De With C, Bezemer P B, Smits G
Clin Chim Acta. 1976 Sep 20;71(3):371-8. doi: 10.1016/0009-8981(76)90087-5.
The Vitatron has been used as designed by the manufacturer under routine laboratory conditions (Broughton, P.M.G., Buttolph, M.A., Gowenlock, A.H., Neill, D.W. and Sleutelberg, R.G. (1969) J. Clin. Pathol. 22, 278). We assessed the possibilities of the AKES with regard to determination of the activities of three enzymes: alanine transaminase (AIT), aspartate transaminase (AsT) and lactate dehydrogenase (LDH) in serum. Precision, accuracy, carry over and sample-diluent contamination were evaluated. This resulted in recommendations for optimal use in terms of capacity and precision, which were supported by computations on a mathematical model for measuring results.
维塔tron已按照制造商的设计在常规实验室条件下使用(布劳顿,P.M.G.,巴托夫,M.A.,戈温洛克,A.H.,尼尔,D.W.和斯勒特尔伯格,R.G.(1969年)《临床病理学杂志》22卷,278页)。我们评估了AKES在测定血清中三种酶活性方面的可能性:丙氨酸转氨酶(AIT)、天冬氨酸转氨酶(AsT)和乳酸脱氢酶(LDH)。评估了精密度、准确性、残留和样品-稀释剂污染情况。这就得出了关于容量和精密度方面最佳使用的建议,这些建议得到了测量结果数学模型计算的支持。