Suppr超能文献

腹主动脉瘤腔内修复术后腹主动脉瘤囊的模块化组件分离与再灌注:一例报告

Modular component separation and reperfusion of abdominal aortic aneurysm sac after endovascular repair of the abdominal aortic aneurysm: a case report.

作者信息

Maleux G, Rousseau H, Otal P, Colombier D, Glock Y, Joffre F

机构信息

Department of Vascular and Interventional Radiology, CHU Rangeuil Toulouse, France.

出版信息

J Vasc Surg. 1998 Aug;28(2):349-52. doi: 10.1016/s0741-5214(98)70172-7.

Abstract

Transluminal placement of a stent graft in patients with an abdominal aortic aneurysm is a new endovascular technique that offers a potentially less invasive and less risky alternative to open surgery. Complications after stent graft placement are not infrequent, but in most cases secondary endovascular intervention is successful. We describe a late major leak in the aneurysmal sac caused by a distal migration of the iliac limb of a bifurcated graft. This late complication was successfully treated by covered stent placement, excluding and thrombosing completely the reformed aneurysm.

摘要

在腹主动脉瘤患者中经腔隙植入覆膜支架是一种新的血管内技术,它为开放手术提供了一种潜在的侵入性较小且风险较低的替代方案。覆膜支架植入后的并发症并不罕见,但在大多数情况下,二次血管内介入治疗是成功的。我们描述了一例由分叉型移植物髂支远端移位导致的动脉瘤囊晚期大渗漏。通过植入覆膜支架成功治疗了这一晚期并发症,完全排除并使重新形成的动脉瘤血栓形成。