Noninvasive estimation of serum bilirubin.

作者信息

Tayaba R, Gribetz D, Gribetz I, Holzman I R

机构信息

Department of Pediatrics, City Hospital Center at Elmhurst, Queens, New York, USA.

出版信息

Pediatrics. 1998 Sep;102(3):E28. doi: 10.1542/peds.102.3.e28.

Abstract

OBJECTIVE

The aim of this study was to evaluate the accuracy and clinical usefulness of a new computer-driven, hand-held device (Chromatics Colormate III) to estimate serum bilirubin from skin-reflectance (skin color) of neonates.

STUDY DESIGN

A total of 2441 infants (both term and premature) at two hospitals had repeated measurements of transcutaneous bilirubin. Of these infants, 900 had one or more laboratory determinations of serum bilirubin. Many of the infants had early measurements of skin color before the onset of jaundice. A visual estimate of the degree of jaundice was made by a health care worker when the laboratory study was drawn. A subgroup of 61 infants was also studied while undergoing phototherapy with a total of 284 comparisons obtained. The reproducibility of the instrument was assessed separately using standardized color tiles and repeated measurements by multiple operators.

RESULTS

The range of serum bilirubin measurements that had concurrent skin color measurements was 3.22 to 338.1 micromol/L (0.2 to 21 mg/dL). The linear regression indicated an r = 0.956, and 95% of the skin color measurements were within 32.2 micromol/L (2.0 mg/dL). There was no interference with the accuracy of the device because of infant race or weight, or because of the use of phototherapy. The device provided reproducible information when infants were tested repeatedly over 30 minutes; the coefficient of variation for the transcutaneous bilirubin measurement was 3.1% around a mean estimate of 135.32 micromol/L (8.4 mg/dL).

CONCLUSION

The Chromatics Colormate III allows for a clinically useful estimate of serum bilirubin in a wide variety of infants. By using a color discrimination algorithm and obtaining a skin measurement before the onset of icterus, this instrument can provide valuable clinical information that obviates the need for serum bilirubin determinations. Its use in newborn nurseries may allow physicians to shorten length of stay more safely and decrease the use of invasive blood tests.

摘要

目的

本研究旨在评估一种新型计算机驱动的手持式设备(Chromatics Colormate III)根据新生儿皮肤反射率(肤色)估算血清胆红素的准确性和临床实用性。

研究设计

两家医院的2441名婴儿(足月儿和早产儿)接受了经皮胆红素的重复测量。其中900名婴儿进行了一次或多次血清胆红素的实验室测定。许多婴儿在黄疸出现之前就进行了早期肤色测量。在进行实验室检查时,由医护人员对黄疸程度进行视觉评估。还对61名接受光疗的婴儿进行了亚组研究,共获得284次比较结果。使用标准化色板并由多名操作人员进行重复测量,分别评估了该仪器的重复性。

结果

同时进行肤色测量的血清胆红素测量范围为3.22至338.1微摩尔/升(0.2至21毫克/分升)。线性回归显示r = 0.956,95%的肤色测量值在32.2微摩尔/升(2.0毫克/分升)范围内。婴儿种族、体重或光疗的使用均未干扰该设备的准确性。当在30分钟内对婴儿进行重复测试时,该设备可提供可重复的信息;经皮胆红素测量的变异系数在平均估计值135.32微摩尔/升(8.4毫克/分升)左右为3.1%。

结论

Chromatics Colormate III能够对各类婴儿的血清胆红素进行具有临床实用性的估算。通过使用颜色辨别算法并在黄疸出现之前进行皮肤测量,该仪器可提供有价值的临床信息,从而无需进行血清胆红素测定。在新生儿病房使用该仪器可能会使医生更安全地缩短住院时间,并减少侵入性血液检查的使用。