Pang G, Luhta R, Rowlands J A
Department of Medical Biophysics, University of Toronto, Sunnybrook Health Science Center, Canada.
Med Phys. 1998 Aug;25(8):1533-46. doi: 10.1118/1.598328.
Large-area x-ray sensitive vidicons have the potential to be superior to conventional x-ray image intensifiers for medical fluoroscopy. To build a large-area x-ray vidicon, an electron lens system is necessary to provide perpendicular beam landing. In this paper, we discuss theoretically how to build two and three cylinder coaxial lenses for this purpose. Some general design rules and a preferred design which satisfies our design criteria are presented. Limitations of these simple types of electron lens are discussed. Two advanced designs for vidicons with short lengths or reduced beam landing error are also included. Among them, the deflection-center-change system is introduced here for the first time.
大面积X射线敏感视像管在医学荧光检查方面有可能优于传统的X射线影像增强器。要制造大面积X射线视像管,需要一个电子透镜系统来实现电子束垂直着陆。在本文中,我们从理论上讨论了如何为此目的构建双圆筒和三圆筒同轴透镜。给出了一些通用设计规则以及一种满足我们设计标准的优选设计。讨论了这些简单类型电子透镜的局限性。还包括了两种针对短长度视像管或降低电子束着陆误差的先进设计。其中,偏转中心改变系统在此首次被引入。