Buchholz T A, Bilton S, Gurgoze E, Strom E A, McNeese M D, Bice W S, Prestidge B R
Department of Radiation Oncology, M.D. Anderson Cancer Center, Houston, Texas 77030, USA.
Radiat Oncol Investig. 1998;6(4):191-8. doi: 10.1002/(SICI)1520-6823(1998)6:4<191::AID-ROI7>3.0.CO;2-T.
The purpose of this paper is to determine whether using off-axis isoseparation curves to optimize the collimator rotation angle improves dose homogeneity. Eleven intact breast irradiation patients underwent computerized tomography (CT) treatment planning with 1 cm abutting slices. Central plane treatment planning, using 6 MV photons, tissue inhomogeneity corrections, and isocentric opposed tangent treatment fields, was performed. Collimators were rotated to match chest wall slope through the use of a beam's-eye-view setting. Patient separations were measured from the apex of the breast to the posterior field border on each axial CT slice. Sagittal-plane isoseparation curves were constructed from these measurements. Using these curves, the collimator rotation that minimized off-axis separation differences was determined. A comparison of off-axis dose inhomogeneity was performed for patients with a > or =10 degrees difference between this optimized collimator angle and the angle determined by chest wall slope. These comparative treatment plans differed only with respect to collimator angle rotation. The mean optimal collimator rotation angle differed significantly from the mean rotation angle which matched the chest wall slope (5.4 degrees vs. 11 degrees, respectively, P < 0.001). Four of the 11 patients had rotation angle differences of 10 degrees. In these patients, the optimization of collimator angle reduced the percentage of breast volume to "that" received > or =110% of the prescribed dose. For the patient with the largest breast size to the patient with the smallest breast size the decreases were, respectively, 5% (15% to 10%), 3% (24% to 21%), 1% (4% to 3%), and 1% (0.9% to 0%). The mean reduction in dose inhomogeneity was greatest in the inferior breast quadrants. At 6 cm and 4 cm off axis, the mean reductions in the percentages of the breast tissue to "that" received 110% of the prescribed dose were respectively 15.1% and 5.3 %. Optimizing the collimator angle through the use of isoseparation curves decreases dose inhomogeneity. The greatest improvements are in the inferior quadrants of the intact breast. The improved dose homogeneity may have clinical relevance in the treatment of patients with large breast sizes.
本文的目的是确定使用离轴等距曲线优化准直器旋转角度是否能改善剂量均匀性。11例完整乳房照射患者接受了层厚1 cm的计算机断层扫描(CT)治疗计划。采用6 MV光子进行中心平面治疗计划,进行组织不均匀性校正,并采用等中心对穿切线野。通过射野视角设置旋转准直器以匹配胸壁斜率。在每个轴向CT切片上测量从乳房顶点到后野边界的患者间距。根据这些测量结果构建矢状面等距曲线。利用这些曲线,确定使离轴间距差异最小化的准直器旋转角度。对于优化后的准直器角度与根据胸壁斜率确定的角度相差≥10°的患者,进行离轴剂量不均匀性比较。这些对比治疗计划仅在准直器角度旋转方面有所不同。平均最佳准直器旋转角度与匹配胸壁斜率的平均旋转角度有显著差异(分别为5.4°和11°,P<0.001)。11例患者中有4例旋转角度差异为10°。在这些患者中,准直器角度的优化降低了接受≥规定剂量110%的乳房体积百分比。从乳房最大的患者到乳房最小的患者,降低幅度分别为5%(从15%降至10%)、3%(从24%降至21%)、1%(从4%降至3%)和1%(从0.9%降至0%)。剂量不均匀性的平均降低在乳房下象限最为显著。在离轴6 cm和4 cm处,接受规定剂量110%的乳房组织百分比的平均降低分别为15.1%和5.3%。通过使用等距曲线优化准直器角度可降低剂量不均匀性。最大的改善在完整乳房的下象限。改善后的剂量均匀性在大乳房患者的治疗中可能具有临床意义。