Suppr超能文献

一种用于人体受试者的快速响应式直接和间接热量计组合装置。

A fast responding combined direct and indirect calorimeter for human subjects.

作者信息

Faber P, Lammert O, Johansen O, Garby L

机构信息

Department of Physical Education, Odense University, Denmark.

出版信息

Med Eng Phys. 1998 Jun;20(4):291-301. doi: 10.1016/s1350-4533(98)00037-x.

Abstract

The construction and performance of a 5.4 m3 combined direct and indirect calorimeter for human subjects is described. The calorimeter was constructed for studies on human subjects primarily undergoing fast alterations in heat production and heat losses, e.g. after a meal or during physical exercise. A heat sink and a heat substitution principle is used to measure sensible heat losses directly. Evaporative heat losses are determined by measuring water vapour input and output. Indirect calorimetry is performed by measuring output air flow and changes in gas composition of the air entering and leaving the calorimeter. The response times (90%) for sensible heat and evaporative heat were found to be 4 min and 34 min, respectively. Three alcohol combustion tests gave a recovery of CO2 and O2 in the range 92-97%. The recovery of water was found to be in the range 56-89%.

摘要

本文描述了一种用于人体受试者的5.4立方米直接和间接组合式热量计的构建与性能。该热量计主要用于研究人体受试者在热量产生和热量损失快速变化的情况下,例如进食后或体育锻炼期间。通过散热器和热替代原理直接测量显热损失。通过测量水蒸气的输入和输出确定蒸发热损失。间接量热法通过测量输出气流以及进入和离开热量计的空气的气体成分变化来进行。显热和蒸发热的响应时间(90%)分别为4分钟和34分钟。三次酒精燃烧测试中二氧化碳和氧气的回收率在92%-97%范围内。水的回收率在56%-89%范围内。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验