Vahlensieck M, Gieseke J, Träber F, Schild H
Radiologische Universitätsklinik Bonn.
Rofo. 1998 Aug;169(2):195-7. doi: 10.1055/s-2007-1015073.
This study aimed at evaluating MTC-sensitivity of frequently used MR-sequences.
7 sequences (T1-, T2-, PD-weighted SE, four gradient-echo sequences) were combined with a 4 element on-resonance MT-Pulse at 0.5 Tesla (T5, Philips medical systems). The measurements were performed at a knee joint and MT-ratios of cartilage, muscle, fat and a copper sulfate phantom were calculated.
Proton density weighted gradient and to a lesser degree spin-echo sequences were superior to T1- and T2-weighted sequences in visualising the MT effects.
It is important to choose an appropriate MR-sequence for MTC experiments. Proton-density gradient echo sequences are recommended for that purpose. T1- and T2-weighted SE sequences should be avoided.
本研究旨在评估常用磁共振序列的磁转移对比(MTC)敏感性。
7种序列(T1加权、T2加权、质子密度加权自旋回波序列,4种梯度回波序列)与一个0.5特斯拉的4元素共振磁转移脉冲(T5,飞利浦医疗系统)相结合。测量在膝关节进行,并计算软骨、肌肉、脂肪和硫酸铜体模的磁转移率。
质子密度加权梯度回波序列以及在较小程度上自旋回波序列在显示磁转移效应方面优于T1加权和T2加权序列。
为磁转移对比实验选择合适的磁共振序列很重要。为此推荐使用质子密度梯度回波序列。应避免使用T1加权和T2加权自旋回波序列。