Evaluation of five devices for self-monitoring of blood glucose in the normoglycaemic range.

作者信息

Glasmacher A G, Brennemann W, Hahn C, Marko C, Wehnert S, Löhr U, Kutschkow R

机构信息

Department of General Internal Medicine, Rheinische Friederich-Wilhelms-Universität, Bonn, Germany.

出版信息

Exp Clin Endocrinol Diabetes. 1998;106(4):360-4. doi: 10.1055/s-0029-1211998.

Abstract

Although hypoglycaemia is a serious risk of intensive insulin therapy, studies on the accuracy of devices for self-monitoring of blood glucose (SMBG) include only a few values below 100 mg/dl. This study evaluated five SMBG devices in the normal range of blood glucose. In three separate series, capillary blood was taken from 207, 214 and 49 healthy subjects and was analysed with the SMBG devices and a laboratory method. Mean (+/- standard deviation) blood glucose values were 86+/-13 mg/dl, 83+/-12 mg/dl and 90+/-16 in series 1-3. All measurements with SMBG devices differed significantly from the laboratory method. The relative deviation from the laboratory method was 5.5% (2.4-10.5%, median and interquartile range) for Accutrend, 7.1% (3.6-12.6%) for Precision QID, 7.5% (3.7-13.7%) for Accu-Chek III, 14.1% (6.2-22.2%) for Companion 2 and 15.2% (10.3-21.3%) for One Touch II (p < 0.05). In conclusion, this study shows important differences in the accuracy of five SMBG devices in the normoglycaemic range. Accutrend and--to a lesser degree--Accu-Chek III and Precision QID obtained better results and can be recommended for SMBG in patients treated with intensive insulin therapy.

摘要

尽管低血糖是强化胰岛素治疗的严重风险,但关于自我血糖监测(SMBG)设备准确性的研究仅纳入了少数低于100mg/dl的值。本研究评估了五种在正常血糖范围内的SMBG设备。在三个独立系列中,从207名、214名和49名健康受试者采集毛细血管血,并用SMBG设备和实验室方法进行分析。系列1-3中平均(±标准差)血糖值分别为86±13mg/dl、83±12mg/dl和90±16mg/dl。使用SMBG设备的所有测量结果与实验室方法均有显著差异。与实验室方法相比,Accutrend的相对偏差为5.5%(2.4-10.5%,中位数和四分位数间距),Precision QID为7.1%(3.6-12.6%),Accu-Chek III为7.5%(3.7-13.7%),Companion 2为14.1%(6.2-22.2%),One Touch II为15.2%(10.3-21.3%)(p<0.05)。总之,本研究表明五种SMBG设备在正常血糖范围内的准确性存在重要差异。Accutrend以及在较小程度上Accu-Chek III和Precision QID取得了更好的结果,可推荐用于强化胰岛素治疗患者的SMBG。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验