Brambilla E, Felloni A, Fadini B L, Strohmenger C L
Department of Pedodontics, S. Paolo Biomedical Sciences Institute, University of Milan, Italy.
Arch Oral Biol. 1998 Oct;43(10):819-23. doi: 10.1016/s0003-9969(98)00060-0.
A simplified method for the determination ionic fluoride (F-) concentrations in microvolume samples using a F(-)-selective electrode is described. Portions (20 microl) of a series of F- standard solutions (1000-0.4 parts/10(9) F-) were placed between the surface of the F- electrode and the surface of a reference electrode mounted end to end with the aid of a steel holder. In this position the two surfaces were used as an analysis chamber, retaining the sample by surface tension without the use of sleeves or other devices. The precision of the results and the absolute detection limit of the modified method were determined. The analytical performance of this set-up was tested by comparing the results with those obtained from the conventional method. The regression between the modified-method data and the conventional-technique data showed a strong correlation (R2 = 0.993) between the results obtained using the two techniques. In the range of F- concentrations from 1000 and 0.48 parts/10(9), the precision of the results (coefficient of variation) was of 5-6%, improving to 0.9% at 0.9 parts/10(9) level.
描述了一种使用氟离子选择性电极测定微量样品中离子氟(F-)浓度的简化方法。借助钢支架,将一系列氟标准溶液(1000 - 0.4 份/10⁹ F-)的部分(20 微升)置于氟电极表面与端对端安装的参比电极表面之间。在此位置,这两个表面用作分析室,通过表面张力保留样品,无需使用套管或其他装置。确定了改进方法的结果精度和绝对检测限。通过将结果与传统方法获得的结果进行比较,测试了该装置的分析性能。改进方法数据与传统技术数据之间的回归显示,两种技术获得的结果之间具有很强的相关性(R² = 0.993)。在 1000 至 0.48 份/10⁹ 的氟浓度范围内,结果精度(变异系数)为 5 - 6%,在 0.9 份/10⁹ 水平时提高到 0.9%。