Ekstrand J
Calcif Tissue Res. 1977 Oct 20;23(3):225-8. doi: 10.1007/BF02012789.
A simple and accurate technique for the determination of fluoride (F-) in capillary-sampled blood is presented. The method is based on the known addition-slope determination technique using the fluoride electrode. No calibration curve or calibration of the electrode is required. A standard deviation of 1.3-5.6% in the range 300-10 ng F/ml was given by 259 duplicate determinations on human plasma. Measurements of parotid saliva showed that it was possible to analayze fluoride concentrations as low as 4 ng F/ml with good reproducibility. The results also show that a 150 microliter sample is sufficient for fluoride determination with high accurary. This method can easily be used in clinical laboratory.
本文介绍了一种用于测定毛细管采集血液中氟化物(F-)的简单准确技术。该方法基于使用氟电极的已知加样斜率测定技术。无需校准曲线或电极校准。对人血浆进行的259次重复测定表明,在300 - 10 ng F/ml范围内,标准偏差为1.3 - 5.6%。腮腺唾液的测量结果表明,能够分析低至4 ng F/ml的氟化物浓度,且重现性良好。结果还表明,150微升样品足以高精度地测定氟化物。该方法可轻松应用于临床实验室。