A new method for combined isometric and isobaric pharmacodynamic studies on porcine coronary arteries.

作者信息

Tankó L B, Mikkelsen E O, Frøbert O, Bagger J P, Gregersen H

机构信息

Institute of Pharmacology, Aarhus University, Denmark.

出版信息

Clin Exp Pharmacol Physiol. 1998 Nov;25(11):919-27. doi: 10.1111/j.1440-1681.1998.tb02344.x.

Abstract
  1. The principal aim of the present study was to explore the isometric and isobaric capacity of a new intravascular technique, impedance planimetry, in basic pharmacodynamic investigations on porcine isolated epicardial coronary arteries. 2. The balloon-based catheter technique provides simultaneous measurements of luminal cross-sectional area and pressure. Sources of errors that may influence the accuracy of measurements were evaluated in detail. 3. Under isometric conditions, the stretch ratio-tension diagram showed typical developments of resting and active tensions of the smooth muscle when exposed to alternating maximal K+ depolarization and mechanical stretching. The mean (+/- SEM) maximum active tension was 28.43 +/- 1.72 mN/mm, which was reached at a stretch ratio of 1.26 +/- 0.02, corresponding to a resting tension of 10.50 +/- 0.53 mN/mm (n = 7). The concentration-response relationship to K+ at optimal basal tension was characterized by a mean (+/- SEM) pD2 value of 1.67 +/- 0.01 (n = 7). 4. Under isobaric conditions in the pressure range 40-140 mmHg, the method allowed the investigation of active vascular responses to partial K+ depolarization. The maximal active response to 25 mmol/L K+ was found at the transmural pressure of 60 mmHg (n = 7). To obtain full K+ concentration-response curves, a basal tension corresponding to a transmural pressure of 120 mmHg was required. The mean (+/- SEM) pD2 value for the concentration-response relationship to K+ was 1.53 +/- 0.01 (n = 10). 5. The vascular sensitivities to cumulatively added K+ and various agonists, such as acetylcholine, 5-hydroxytryptamine and noradrenaline, obtained from the same vessel segment at the same initial conditions corresponding to 120 mmHg were significantly higher with the isometric than with the isobaric approach. 6. The results of the present study suggest that impedance planimetry could be a useful tool in pharmacological and physiological investigations of medium-sized arteries, both under isometric and isobaric conditions.
摘要
  1. 本研究的主要目的是在猪离体心外膜冠状动脉的基础药效学研究中,探索一种新的血管内技术——阻抗平面测量法的等长和等压容量。2. 基于球囊的导管技术可同时测量管腔横截面积和压力。详细评估了可能影响测量准确性的误差来源。3. 在等长条件下,拉伸比 - 张力图显示了平滑肌在交替进行最大钾离子去极化和机械拉伸时静息张力和主动张力的典型变化。平均(±标准误)最大主动张力为28.43±1.72 mN/mm,在拉伸比为1.26±0.02时达到,对应静息张力为10.50±0.53 mN/mm(n = 7)。在最佳基础张力下,对钾离子的浓度 - 反应关系的平均(±标准误)pD2值为1.67±0.01(n = 7)。4. 在40 - 140 mmHg压力范围内的等压条件下,该方法可用于研究血管对部分钾离子去极化的主动反应。在60 mmHg的跨壁压力下发现对25 mmol/L钾离子的最大主动反应(n = 7)。为获得完整的钾离子浓度 - 反应曲线,需要对应120 mmHg跨壁压力的基础张力。对钾离子浓度 - 反应关系的平均(±标准误)pD2值为1.53±0.01(n = 10)。5. 在对应120 mmHg的相同初始条件下,从同一血管段获得的血管对累积添加的钾离子和各种激动剂(如乙酰胆碱、5 - 羟色胺和去甲肾上腺素)的敏感性,等长法显著高于等压法。6. 本研究结果表明,阻抗平面测量法在等长和等压条件下,可能是中动脉药理和生理研究中的一种有用工具。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验