The effect of dose on the characterization of aerodynamic particle-size distributions of beclomethasone dipropionate metered-dose inhalers.

作者信息

el-Araud K A, Clark B J, Kaahwa C, Anum P, Chrystyn H

机构信息

Postgraduate Studies in Pharmaceutical Chemistry, School of Pharmacy, University of Bradford, UK.

出版信息

J Pharm Pharmacol. 1998 Oct;50(10):1081-5. doi: 10.1111/j.2042-7158.1998.tb03316.x.

Abstract

The aerodynamic particle-size distribution for two doses of a Becloforte metered-dose inhaler (MDI) was measured by use of a twin-stage impinger (TSI), the new multi-stage (five-stage) liquid impinger (MSLI) and the Andersen cascade impactor (ACI) (n = 5 for each apparatus). The mean (s.d.) fine-particle doses measured by the three techniques for the Becloforte MDI were 40.3 (1.2), 45.7 (0.5) and 41.8 (0.4)% w/w, respectively; the median mass aerodynamic diameters (MMAD) measured using the MSLI and the ACI were 3.50 and 3.73 microns, respectively. The MSLI fine particle (< 6.8 microns) doses for 2, 5, 10, 20, 30 and 40 doses from Becloforte MDIs (n = 5 for each dose) were 49.7 (0.7), 52.9 (1.2), 45.3 (0.6), 45.5 (0.71), 45.9 (0.7) and 46.4 (0.7)% w/w, respectively. Values obtained using the ACI (< 5.8 microns) were 40.8 (1.0), 41.0 (0.8), 44.4 (0.5), 43.1 (0.4), 42.8 (0.5) and 40.4 (0.4)% w/w (n = 4). MMAD values measured with the MSLI were 3.39, 3.46, 3.75, 3.91, 4.15 and 4.45 microns, respectively; using the ACI they were 3.46, 3.54, 3.61, 3.66, 3.73 and 3.85 microns. The results indicate that the measured aerodynamic particle-size distributions of beclomethasone dipropionate MDIs are affected by the dose dispensed and by the apparatus used for measurement.

摘要

使用双级冲击器(TSI)、新型多级(五级)液体冲击器(MSLI)和安德森级联冲击器(ACI)(每种仪器n = 5)测量了两剂倍氯米松定量吸入器(MDI)的空气动力学粒径分布。通过这三种技术测量的倍氯米松MDI的平均(标准差)细颗粒剂量分别为40.3(1.2)、45.7(0.5)和41.8(0.4)% w/w;使用MSLI和ACI测量的中位质量空气动力学直径(MMAD)分别为3.50和3.73微米。来自倍氯米松MDI的2、5、10、20、30和40剂的MSLI细颗粒(< 6.8微米)剂量分别为49.7(0.7)、52.9(1.2)、45.3(0.6)、45.5(0.71)、45.9(0.7)和46.4(0.7)% w/w。使用ACI(< 5.8微米)获得的值为40.8(1.0)、41.0(0.8)、44.4(0.5)、43.1(0.4)、42.8(0.5)和40.4(0.4)% w/w(n = 4)。使用MSLI测量的MMAD值分别为3.39、3.46、3.75、3.91、4.15和4.45微米;使用ACI测量的MMAD值分别为3.46、3.54、3.61、3.66、3.73和3.85微米。结果表明,所测量的二丙酸倍氯米松MDI的空气动力学粒径分布受所分配剂量和用于测量的仪器的影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验