Suppr超能文献

Preliminary observations on the Colibri CO2-indicator.

作者信息

Petroianu G A, Maleck W H, Bergler W F, Altmannsberger S, Rüfer R

机构信息

Institute of Pharmacology, University of Heidelberg at Mannheim, Germany.

出版信息

Am J Emerg Med. 1998 Nov;16(7):677-80. doi: 10.1016/s0735-6757(98)90175-9.

Abstract

The performance of a new colorimetric CO2-indicator (Colibri) was assessed in mini-pigs. It performed well during 8-hour procedures. Neither nitrous oxide, nor halothane, nor carbon monoxide, nor intratracheal application of drugs (epinephrine, atropine, lidocaine, and naloxone) interfered with its function. It gave a distinct color change at high ventilation frequencies up to 120/min. The only problem observed was difficulty in matching the colors displayed with the comparison color chart provided. The Colibri's performance seems at least equal to that of the EasyCAP detector, although both devices share some disadvantages (no alarms, semiquantitative, difficult reading in the dark). After initial control of endotracheal tube position by an esophageal detector device, both the Colibri and the EasyCAP seem suited for monitoring of ventilation and circulation if quantitative capnometry is unavailable.