Validation of the conductance catheter method for measurement of ventricular volumes under varying conditions relevant to cardiac surgery.

作者信息

Al-Khalidi A H, Townend J N, Bonser R S, Coote J H

机构信息

Department of Physiology, The Medical School, The University of Birmingham, United Kingdom.

出版信息

Am J Cardiol. 1998 Nov 15;82(10):1248-52. doi: 10.1016/s0002-9149(98)00613-4.

Abstract

The development of the conductance catheter method has enabled continuous measurement of intraventricular volume in vivo, thus making assessment of pump performance of the heart in vivo possible using pressure-volume analysis. However, this method has not been validated under conditions where pump rate, conductance, viscosity, and temperature of the fluid in the heart chamber is changed as happens in cardiac surgery. To validate the method, pressure-volume data were measured by conductance catheter in a physical model of the human left ventricle. The volume, salinity, viscosity, and temperature of the fluid inside the model were rigorously controlled. The measured pressure-volume data were compared with the actual values to assess the accuracy and dependence of the conductance-measured volumes on salinity, viscosity, temperature, and pump rate. Conductance-measured volumes were not significantly different over a range of heart rates extending from 60 to 100 beats/min, and they were not significantly different over a salinity range of 0.2 to 2 normal saline, a viscosity range of 2.7 to 3.5 centipoise, or over a temperature range of 20 degrees C to 39 degrees C. The percentage errors between actual stroke volumes and conductance-derived volumes were of the order of 10% on average. Our data shows that for a broad range of conditions normally encountered during cardiac surgery, there is no dependence of conductance-measured volume on heart rate, viscosity, temperature, or salinity, provided the correct value of fluid resistivity is used.

摘要

电导导管法的发展使得在体连续测量心室内容积成为可能,从而利用压力-容积分析在体评估心脏的泵血性能。然而,在心脏手术中出现的泵血速率、电导、黏度和心腔内液体温度发生变化的情况下,该方法尚未得到验证。为了验证该方法,通过电导导管在人体左心室的物理模型中测量压力-容积数据。对模型内液体的容积、盐度、黏度和温度进行了严格控制。将测得的压力-容积数据与实际值进行比较,以评估电导测量容积对盐度、黏度、温度和泵血速率的准确性和依赖性。在心率范围从60次/分钟到100次/分钟的情况下,电导测量的容积没有显著差异,在盐度范围为0.2至2倍生理盐水、黏度范围为2.7至3.5厘泊或温度范围为20摄氏度至39摄氏度的情况下,也没有显著差异。实际每搏输出量与电导得出的容积之间的百分比误差平均约为10%。我们的数据表明,在心脏手术中通常遇到的广泛条件下,只要使用正确的液体电阻率值,电导测量的容积就不依赖于心率、黏度、温度或盐度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验