Braza A J, Modamio P, Mariño E L
Department of Pharmacy and Pharmaceutical Technology, Faculty of Pharmacy, University of Barcelona, Spain.
J Chromatogr B Biomed Sci Appl. 1998 Nov 6;718(2):267-72. doi: 10.1016/s0378-4347(98)00378-8.
Two reversed-phase HPLC methods with UV detection to quantify celiprolol and oxprenolol in human plasma are described. The analytical methods for the determination of both drugs used the same reversed-phase HPLC column, mobile phase and extraction procedure. Linearity was obtained in the ranges 15.63-1000 and 25-800 ng/ml for celiprolol and oxprenolol, respectively. Intra-day and inter-day variation was lower than 14%. After validation of the methods, analytical error functions were established as S.D. (ng/ml)=3.096+0.041C for celiprolol and S.D. (ng/ml)=8.906+8.075x10(-8)C3 for oxprenolol.