Suppr超能文献

Evaluation of the diagnostic performance of the Boehringer Mannheim CEDIA LSD assay.

作者信息

Verstraete A G, Steyaert S

机构信息

Laboratory of Clinical Biology-Toxicology, University Hospital, Gent, Belgium.

出版信息

J Anal Toxicol. 1998 Nov-Dec;22(7):601-4. doi: 10.1093/jat/22.7.601.

Abstract

The precision and the diagnostic performance of the Boehringer Mannheim CEDIA DAU LSD assay was evaluated. The assay was performed in the semi-quantitative mode on a Hitachi 917 analyzer. Within-run coefficients of variation (CVs) of the semiquantitative values for 0.25 and 1.0 ng/mL were 11.2 and 6.2%, respectively. Day-to-day CVs for the same concentrations were 12.6 and 8.6%. We analyzed 318 urine samples by CEDIA, DPC Coat-A-Count RIA and Behring EMIT II. Confirmation was performed by GC-MS, after extraction on Bond Elut Certify columns. Two hundred sixty-three samples were negative by all methods. Twenty-five samples were positive by all immunoassays, 19 of which were confirmed by gas chromatography-mass spectrometry (GC-MS). One sample was falsely negative by CEDIA. Three samples were positive by EMIT and CEDIA, but negative by RIA and GC-MS. Twenty-six samples were positive by EMIT alone, but they were not confirmed by GC-MS. The LSD CEDIA assay seems to be less specific than DPC RIA but more specific than the EMIT LSD assay.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验